483
Cirtec MedicalFDA 483 - Cirtec Medical - March 08, 2023
Record Details
Cirtec Medical in Los Gatos, CA, a contract manufacturer, was cited for inadequate corrective and preventive action procedures. The inspection revealed that the firm failed to open a CAPA to investigate and correct discrepancies in heat sealer calibration parameters. This lapse affected multiple lots of product manufactured prior to the correction, indicating a significant control deficiency.