> China NMPA product recall for Image archiving and transmission system by GE Healthcare (China) Co., Ltd. published September 07, 2015. Recall level: . The National Medical Products Administration (NMPA) issued a notice on September 7, 2015, regarding 

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: GE Healthcare recalls image archiving and transmission systems
- Company Name: GE Healthcare (China) Co., Ltd.
- Publication Date: 2015-09-07
- Product Name: Image archiving and transmission system
- Recall Reason: GE Healthcare recently discovered that using Universal Viewer (versions 5.0 to 5.0 SP7) from the Centricity PACS-IW backend, when the system parameter "MapRoute" is set to a value greater than 1, may result in not all images being processed during the examination, thus missing one or more images. No patient injuries have been reported as a result.
- Discovering Company: GE Healthcare Systems Trade Development (Shanghai) Co., Ltd.
- Manufacturing Company: GE Healthcare (China) Co., Ltd.
- Summary: The National Medical Products Administration (NMPA) issued a notice on September 7, 2015, regarding a voluntary recall initiated by GE Healthcare for its Image Archiving and Transmission Systems. GE Healthcare Trading Development (Shanghai) Co., Ltd. reported the recall of specific models, including the Centricity Universal Viewer, Version 5.0 (and earlier patches). The primary issue identified was a software defect where, if the system parameter "MapRoute" was set above 1, the system might fail to process all images during an examination, leading to missing images. Fortunately, no patient injuries related to this defect have been reported. This recall falls under the regulatory purview of the NMPA in China. A key aspect of this recall is that, despite the product holding a CFDA (now NMPA) registration certificate, all affected batches of the Image Archiving and Transmission Systems were not sold within the Chinese market, with distribution concentrated in regions such as the United States and the European Economic Area. Consequently, no direct corrective actions were necessary for products within China. However, the NMPA requested that all relevant provincial and municipal food and drug administrations enhance their oversight and management of similar medical imaging products. The initial recall report was submitted on August 11, 2015.