# China NMPA Product Recall - treadmill

Source: https://www.globalkeysolutions.net/records/china_product_recall/ge-medical-systems-information-technologies/6ad655c4-b01e-4fa7-b9f3-07d664ada00a/
Source feed: China

> China NMPA product recall for treadmill by GE Medical Systems Information Technologies published March 11, 2015. Recall level: Level II. GE Healthcare Systems Trading Development (Shanghai) Co., Ltd. initiated a voluntary Class II recall

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: GE Healthcare Systems Trade Development (Shanghai) Co., Ltd. voluntarily recalls fortifications.
- Company Name: GE Medical Systems Information Technologies
- Publication Date: 2015-03-11
- Product Name: treadmill
- Recall Level: Level II
- Recall Reason: An internal quality inspection by GE Healthcare revealed that the power cord connected to the T2100 may not have been assembled correctly. If the power cord is not assembled properly and a secondary electrical fault occurs simultaneously (e.g., a worn power cord or the power cord coming into contact with the treadmill chassis), it could potentially cause an electric shock to a patient or operator. However, to date, no incidents involving this problem have been reported.
- Discovering Company: GE Healthcare Systems Trade Development (Shanghai) Co., Ltd.
- Manufacturing Company: GE Medical Systems Information Technologies
- Summary: GE Healthcare Systems Trading Development (Shanghai) Co., Ltd. initiated a voluntary Class II recall for its T2100 Treadmills, as reported to the National Medical Products Administration (NMPA) on February 12, 2015, and published on March 11, 2015. The recall stems from an internal quality inspection revealing that the power cord of the T2100 product might not have been assembled according to specifications. This manufacturing deviation presents a potential electric shock hazard to patients or operators if the improperly assembled power cord experiences a secondary electrical fault, such as wear or contact with the mount's chassis. While no related incidents have been reported, GE Healthcare is taking proactive measures. The company is issuing customer letters to inform affected users of the issue and to guide them through a self-check procedure. Customers whose products do not pass this self-check are instructed to immediately cease using the device and contact GE Healthcare for support. Furthermore, GE Healthcare is deploying engineers to perform on-site corrections for the affected treadmills. This recall impacts 458 units sold in China, Australia/New Zealand, Europe, the USA, and other regions.

Company: https://www.globalkeysolutions.net/companies/ge-medical-systems-information-technologies/4f9305a8-d284-4eb1-bdd2-f15eede2ad53/