# China NMPA Product Recall - Fetal monitor (Model: Corometrics171/172)

Source: https://www.globalkeysolutions.net/records/china_product_recall/ge-medical-systems-information-technologies/86d1f8cd-8b30-4b58-8d85-90364ff8066b/
Source feed: China

> China NMPA product recall for Fetal monitor (Model: Corometrics171/172) by GE Medical Systems Information Technologies published June 03, 2015. Recall level: . On June 3, 2015, GE Medical Systems Information Technologies initiated a voluntary product recall co

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: GE Medical Systems Information Technologies recalls fetal monitors.
- Company Name: GE Medical Systems Information Technologies
- Publication Date: 2015-06-03
- Product Name: Fetal monitor (Model: Corometrics171/172)
- Recall Reason: The alarm lower limit for low fetal heart rate is set incorrectly.
- Discovering Company: GE Healthcare Systems Trade Development (Shanghai) Co., Ltd.
- Manufacturing Company: GE Medical Systems Information Technologies
- Summary: On June 3, 2015, GE Medical Systems Information Technologies initiated a voluntary product recall concerning its Corometrics 171/172 fetal monitors. The reason for this critical action was an identified incorrect setting of the low fetal heart rate alarm lower limit. Such a malfunction could potentially compromise patient safety by failing to alert medical professionals to dangerously low fetal heart rates, thus delaying necessary interventions. The recall notification was disseminated by GE Medical Systems Trade Development (Shanghai) Co., Ltd., which confirmed that the specific recalled devices, identified by Registration Certificate No.: CFDA (Imported) No. 20042210592, had not been distributed or sold within the Chinese market. This voluntary recall was conducted under the oversight of the National Medical Products Administration (NMPA) of China, demonstrating adherence to regulatory expectations for medical device safety. In response to this incident, the NMPA requested that all municipal food and drug administrations strengthen their supervision and management over similar medical products. This proactive measure aims to ensure robust oversight of medical devices and reinforces the regulatory commitment to public health. The company's prompt action, though for products not sold locally, underscores the global responsibility of medical device manufacturers to address safety concerns promptly.

Company: https://www.globalkeysolutions.net/companies/ge-medical-systems-information-technologies/4f9305a8-d284-4eb1-bdd2-f15eede2ad53/