# China NMPA Product Recall - Continuous blood purification equipment

Source: https://www.globalkeysolutions.net/records/china_product_recall/infomed-s-a/8ba9f7f1-33d2-4f86-b4bd-2419fd461509/
Source feed: China

> China NMPA product recall for Continuous blood purification equipment by Infomed S. A. published October 25, 2018. Recall level: Level 3 Recall. 

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: Guangzhou Delang Medical Equipment Co., Ltd. voluntarily recalls continuous blood purification equipment.
- Company Name: Infomed S. A.
- Publication Date: 2018-10-25
- Product Name: Continuous blood purification equipment
- Recall Level: Level 3 Recall
- Recall Reason: The low-limit alarm setting for venous pressure does not comply with the standards YY 0645-2008 and YZB/SWI 3166-2015; the application of partial electric shock protection symbols does not comply with the requirements of GB9706.1-2007 standard.
- Discovering Company: Guangzhou Delang Medical Equipment Co., Ltd.
- Manufacturing Company: Infomed S. A.

Company: https://www.globalkeysolutions.net/companies/infomed-s-a/c1674d0e-7ebc-4f62-a9d0-eb77db1798e3/