# China NMPA Product Recall - Laparoscopic surgical instruments

Source: https://www.globalkeysolutions.net/records/china_product_recall/microline-surgical-inc/3a9619b8-3bb0-491f-ae4f-33353a424ccb/
Source feed: China

> China NMPA product recall for Laparoscopic surgical instruments by Microline Surgical, Inc. published August 02, 2017. Recall level: Level 3. Microline Surgical, Inc. initiated a voluntary Class III recall of its Laparoscopic Surgical Instrum

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: Microline Surgical, Inc. voluntarily recalls laparoscopic surgical instruments.
- Company Name: Microline Surgical, Inc.
- Publication Date: 2017-08-02
- Product Name: Laparoscopic surgical instruments
- Recall Level: Level 3
- Recall Reason: The insulating layer made of black heat-shrink material on the ReNew repetitive manipulation heads (grippers and separating forceps) has a potential risk of cracking after high-temperature and high-pressure sterilization/re-sterilization.
- Discovering Company: Beijing Weini Huili Medical Equipment Co., Ltd.
- Manufacturing Company: Microline Surgical, Inc.
- Summary: Microline Surgical, Inc. initiated a voluntary Class III recall of its Laparoscopic Surgical Instruments, as reported by Beijing Weini Huili Medical Device Co., Ltd. and regulated by the National Medical Products Administration (NMPA) under CFDA (Imported) 2014 No. 3220762. The recall notice was published on August 2, 2017, stemming from a report dated July 31, 2017.

The core issue involves a potential risk with the insulation layer of the ReNew repetitive manipulation heads (grasping and dissecting forceps). This black heat-shrinkable material was found to be susceptible to cracking after undergoing high-temperature and high-pressure sterilization or re-sterilization processes. This defect could compromise the instrument's safety and functionality during laparoscopic surgical procedures.

While Microline Surgical, Inc. has not received any reports of patient harm or adverse events directly linked to this insulation layer issue, the recall was initiated globally as a precautionary measure to avert potential risks. The recall, classified as Level 3, encompasses various models and all associated batch numbers. Corrective actions included promptly issuing recall notices to domestic customers and ensuring all affected products are returned to the manufacturer for appropriate processing, demonstrating adherence to regulatory requirements and commitment to patient safety.

Company: https://www.globalkeysolutions.net/companies/microline-surgical-inc/cd08eb9f-d203-4b38-b5b4-4f5344957a6e/