> China NMPA product recall for Baby warming station by Ohmeda Medial, A Division of Datex Ohmeda Inc. published December 07, 2016. Recall level: Level II. GE Healthcare Systems Trading Development (Shanghai) Co., Ltd. initiated a voluntary Class II recall

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: GE Healthcare Systems Trade Development (Shanghai) Co., Ltd. has initiated a voluntary recall of baby warmers.
- Company Name: Ohmeda Medial, A Division of Datex Ohmeda Inc.
- Publication Date: 2016-12-07
- Product Name: Baby warming station
- Recall Level: Level II
- Recall Reason: The screws of the heating element in a baby warmer may loosen and fall onto the bed after prolonged use. This poses a potential risk of heat injury to the patient in clinical practice.
- Discovering Company: GE Healthcare Systems Trade Development (Shanghai) Co., Ltd.
- Manufacturing Company: Ohmeda Medial, A Division of Datex Ohmeda Inc.
- Summary: GE Healthcare Systems Trading Development (Shanghai) Co., Ltd. initiated a voluntary Class II recall for its Lullaby model Baby Warmers, as reported to the National Medical Products Administration (NMPA) on October 19, 2016. The manufacturer is Ohmeda Medial, A Division of Datex Ohmeda Inc. The critical issue identified is that screws within the heating element of these devices may loosen after prolonged use, potentially detaching and falling onto the infant's bed, which poses a risk of heat injury to patients. The affected products, bearing Registration Certificate Number CFDA (Imported) No. 20133541372, are designed to provide controlled infrared heat for infants. This recall, referenced under NMPA Index No. JGXX-2016-10396, impacts 270 units sold in China out of 278 imported, with global implications extending to regions like the European Economic Area. As a required action, GE Healthcare will issue a comprehensive customer letter to all users, detailing the safety concern. Additionally, a supplementary appendix will be provided for the service manual, offering specific guidance to maintenance personnel to ensure proper product servicing and mitigate the identified risk, thereby preventing potential patient harm.