# China NMPA Product Recall - Infant incubator, multi-functional incubator, phototherapy equipment, infant warming table, infant radiant warming table

Source: https://www.globalkeysolutions.net/records/china_product_recall/ohmeda-medical/8f8a3f64-4780-4e6a-8404-10e171911ff4/
Source feed: China

> China NMPA product recall for Infant incubator, multi-functional incubator, phototherapy equipment, infant warming table, infant radiant warming table by Ohmeda Medical published January 19, 2017. Recall level: . The National Medical Products Administration (NMPA) published a recall notice on January 19, 2017, c

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: Ohmeda Medical recalls products including infant incubators.
- Company Name: Ohmeda Medical
- Publication Date: 2017-01-19
- Product Name: Infant incubator, multi-functional incubator, phototherapy equipment, infant warming table, infant radiant warming table
- Recall Reason: American-style power cords may overheat, causing fires, charring, smoke, or sparks at the connection between the power cord and the wall socket, resulting in heat damage.
- Discovering Company: GE Healthcare Systems Trade Development (Shanghai) Co., Ltd.
- Manufacturing Company: Ohmeda Medical
- Summary: The National Medical Products Administration (NMPA) published a recall notice on January 19, 2017, concerning medical devices manufactured by Ohmeda Medical, with GE Healthcare Systems Trade Development (Shanghai) Co., Ltd. acting as the responsible unit in China. This voluntary recall, reported on December 21, 2016, addresses a potential safety issue involving specific US-style power cords, known as 'Taller Bridge' cords, manufactured by Electri-Cord. The main concern is the overheating of these power cords, which could lead to fire, burning, smoke, or sparks at the connection point between the cord and the wall outlet, potentially causing heat damage. The recall encompasses various infant care products, including different models of infant incubators, multifunctional incubators, phototherapy equipment, infant warming tables, and infant radiant warming tables. While Ohmeda Medical initiated this recall due to the identified safety hazard, the affected product batches were not distributed or sold within China. Despite this, the NMPA requested provincial, autonomous region, and municipal food and drug administrations to strengthen their supervision and management of similar medical devices. No injuries related to this issue were reported.

Company: https://www.globalkeysolutions.net/companies/ohmeda-medical/1b069dd9-c5ca-4d87-890f-3b3d07bedf7f/