# China NMPA Product Recall - Quick-release handle

Source: https://www.globalkeysolutions.net/records/china_product_recall/shanghai-minimally-invasive-orthopedic-medical-technology-co-ltd/8f7b93d2-16da-4425-b15e-5610aa9c7508/
Source feed: China

> China NMPA product recall for Quick-release handle by Shanghai Minimally Invasive Orthopedic Medical Technology Co., Ltd. published December 15, 2016. Recall level: Level II. Shanghai Minimally Invasive Orthopedic Medical Technology Co., Ltd., also identified as Microport Or

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: Shanghai Minimally Invasive Orthopedic Medical Technology Co., Ltd. is voluntarily recalling the "quick-release handle" in its "hip joint surgical instrument kit".
- Company Name: Shanghai Minimally Invasive Orthopedic Medical Technology Co., Ltd.
- Publication Date: 2016-12-15
- Product Name: Quick-release handle
- Recall Level: Level II
- Recall Reason: 1) The bone file cannot lock properly onto the handle. 2) After forcibly locking the bone chain, the bone file cannot disengage properly from the handle.
- Discovering Company: Shanghai Minimally Invasive Orthopedic Medical Technology Co., Ltd.
- Manufacturing Company: Shanghai Minimally Invasive Orthopedic Medical Technology Co., Ltd.
- Summary: Shanghai Minimally Invasive Orthopedic Medical Technology Co., Ltd., also identified as Microport Orthopedics, Inc., initiated a voluntary Class II recall of its "Quick-Connect Handle," a component of the "Hip Joint Surgical Instrument Kit." This recall, documented under National Medical Products Administration (NMPA) Index No. JGXX-2016-10431 and published on December 15, 2016, stemmed from the company's internal investigation.

The core issues identified involved critical functional defects with the "Quick-Connect Handle," specifically its interaction with the bone reamer. The investigation revealed that the bone reamer could not be properly locked onto the handle. Additionally, if forcibly engaged, the bone reamer subsequently could not be properly disengaged. These malfunctions pose a significant concern for patient safety and surgical efficiency during hip joint procedures.

The required action taken by the company was to recall the affected products. While three units had been imported into China, these were successfully returned to the United States in September 2016. As a result, at the time of the report, no affected batch numbers were present in the Chinese domestic market, and thus, no further direct corrective action regarding existing domestic inventory was required beyond the documented return of the imported units.

Company: https://www.globalkeysolutions.net/companies/shanghai-minimally-invasive-orthopedic-medical-technology-co-ltd/8c61a790-072a-4ccd-b907-8769f72c1fa8/