# China NMPA Product Recall - Disposable suction device

Source: https://www.globalkeysolutions.net/records/china_product_recall/sichuan-hualikang-medical-technology-co-ltd/b489f546-4d54-48b4-84e7-a451def2f269/
Source feed: China

> China NMPA product recall for Disposable suction device by Sichuan Hualikang Medical Technology Co., Ltd. published November 25, 2019. Recall level: Level 3 Recall. Sichuan Hualikang Medical Technology Co., Ltd. has initiated a voluntary Class III recall of its dis

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: Sichuan Hualikang Medical Technology Co., Ltd. voluntarily recalls disposable suction devices.
- Company Name: Sichuan Hualikang Medical Technology Co., Ltd.
- Publication Date: 2019-11-25
- Product Name: Disposable suction device
- Recall Level: Level 3 Recall
- Recall Reason: The product in question failed the national supervision and spot check in 2019. The tested sample showed vacuum residue exceeding the requirements of the YY0339-2009 standard for respiratory products.
- Discovering Company: Sichuan Hualikang Medical Technology Co., Ltd.
- Manufacturing Company: Sichuan Hualikang Medical Technology Co., Ltd.
- Summary: Sichuan Hualikang Medical Technology Co., Ltd. has initiated a voluntary Class III recall of its disposable suction devices. This action follows a national supervision and spot check conducted in 2019, which identified a critical quality issue with the company's products. Specifically, tested samples of the disposable suction devices demonstrated vacuum residue levels exceeding the stringent requirements specified in the YY0339-2009 standard for "Respiratory Suction Devices." This deviation from established performance criteria necessitated the recall, managed under the oversight of the National Medical Products Administration (NMPA). The standard YY0339-2009 is a key regulatory framework governing the safety and efficacy of such medical devices, and exceeding its vacuum residue limits could potentially impact device function or patient safety. The affected products, registered under numbers such as 川蓉械备20170003号, are subject to this recall. Detailed information concerning the specific types and batch numbers of the impacted devices is available in the associated "Medical Device Recall Event Report Form." This proactive recall by Sichuan Hualikang Medical Technology Co., Ltd. highlights a commitment to addressing product non-conformities and upholding the quality and safety standards mandated by the NMPA.

Company: https://www.globalkeysolutions.net/companies/sichuan-hualikang-medical-technology-co-ltd/b82df35f-8f44-4474-afdf-be28a996d237/