# China NMPA Product Recall - Spinal surgical tools

Source: https://www.globalkeysolutions.net/records/china_product_recall/synthes-gmbh/0a24d619-5770-421c-9337-1a7acc7894c7/
Source feed: China

> China NMPA product recall for Spinal surgical tools by Synthes GmbH published October 17, 2013. Recall level: . A National Medical Products Administration (NMPA) notice, published on October 17, 2013, announced a

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: Synthes GmbH recalls spinal surgery tools
- Company Name: Synthes GmbH
- Publication Date: 2013-10-17
- Product Name: Spinal surgical tools
- Recall Reason: Synthes received notification from one of its suppliers that one or more bars in a batch of materials used to produce the affected parts were mixed with the wrong material. The specified material is 420A, and the material mixed in with the bars is X20CrMnNi 12 8 6 (a non-magnetic austenitic steel grade used in generator manufacturing). Using the wrong material will result in the device having lower hardness and corrosion resistance than specified.
- Discovering Company: Johnson & Johnson (Shanghai) Medical Devices Co., Ltd.
- Manufacturing Company: Synthes GmbH
- Summary: A National Medical Products Administration (NMPA) notice, published on October 17, 2013, announced a voluntary recall of spinal surgery instruments by manufacturer Synthes GmbH. The recall was initiated due to a critical manufacturing issue: incorrect material (X20CrMnNi 12 8 6) was inadvertently used in place of the specified 420A material during production. This material discrepancy results in the affected instruments having lower hardness and reduced corrosion resistance, potentially compromising their performance and safety. Johnson & Johnson (Shanghai) Medical Devices Co., Ltd., the responsible entity in China, confirmed this issue. Although the products are registered with NMPA import permits, no affected units were imported or sold in China. Consequently, the recall is an overseas action, primarily impacting Spain, Switzerland, Netherlands, USA, and Canada, for specific product numbers like T987734 and T988931. Johnson & Johnson Shanghai reported the situation to the NMPA but stated no further recall measures would be undertaken domestically. The NMPA advised provincial food and drug administrations to strengthen supervision over such products, emphasizing the critical need for precise material control in medical device manufacturing.