# China NMPA Product Recall - Spinal surgical instruments (brand name: Synthes)

Source: https://www.globalkeysolutions.net/records/china_product_recall/synthes-gmbh/a2f60e18-ea9b-44d8-ba9d-002a2bdaeecd/
Source feed: China

> China NMPA product recall for Spinal surgical instruments (brand name: Synthes) by Synthes GmbH published December 12, 2014. Recall level: Level I. On December 12, 2014, the National Medical Products Administration (NMPA) and the Shanghai Food and 

---

## Details

- Record Type: CHINA_PRODUCT_RECALL
- Title: Johnson & Johnson (Shanghai) Medical Devices Co., Ltd. voluntarily recalls spinal surgery tools.
- Company Name: Synthes GmbH
- Publication Date: 2014-12-12
- Product Name: Spinal surgical instruments (brand name: Synthes)
- Recall Level: Level I
- Recall Reason: The actual length of the tip of the 2.0mm drill bit with a 14mm depth limiter/Zero-P used for spinal surgery is 16mm, which is longer than the stated 14mm. This defective instrument may cause bone injury, soft tissue/intervertebral disc injury, spinal cord/meningeal injury, and surgical delay.
- Discovering Company: Johnson & Johnson (Shanghai) Medical Devices Co., Ltd.
- Manufacturing Company: Synthes GmbH
- Summary: On December 12, 2014, the National Medical Products Administration (NMPA) and the Shanghai Food and Drug Administration announced a Class I voluntary recall initiated by Synthes GmbH, with Johnson & Johnson (Shanghai) Medical Devices Co., Ltd. serving as the responsible unit in China. The recall involves specific spinal surgery tools, a 2.0mm drill bit with a 14mm depth limiter/Zero-P (specification: 03.617.914, batch number: 8496070). The primary issue identified was a manufacturing defect: the drill bit, specified to have a 14mm tip length, actually measured 16mm. This discrepancy was discovered during a surgical procedure, though it did not cause patient harm in that specific instance. However, the use of such a defective instrument carries potential risks including bone injury, soft tissue or intervertebral disc damage, spinal cord or meningeal injury, and surgical delays. Synthes GmbH conducted an investigation and found no similar complaints regarding this model and batch number, either globally or in China. Despite only 6 units being imported to China, with 2 sold, the company is undertaking a global recall. Required actions include notifying affected distributors and instructing them to return all impacted products to the original manufacturer, ensuring adherence to the NMPA's regulatory framework.

Company: https://www.globalkeysolutions.net/companies/synthes-gmbh/77a06946-ba7a-4bf2-916a-600eabfb0ce7/